The ELF laboratory furnaces comprise 3 bench mounted models designed for light duty and general use up to 1100 °C.


• Max. Temp. 1100 °C

• Volume 6 liters

• Drop down door

• Ventilated via top mounted ceramic chimney

• Controller 301 with single ramp to setpoint

 and timer


These laboratory furnaces have a simple drop down door and a top mounted ceramic chimney.

The combination of low thermal mass insulation and free radiating wire elements embedded in the chamber sides provide efficient heating.

Standard features

  • 1100°C maximum operating temperature
  • 6, 14 or 23 litre chamber volumes
  • Drop down door with air gap to minimise external temperature
  • Carbolite 301 controller, with single ramp to set-point & process timer
  • Delayed start / process timer function as standard
  • Vacuum formed, low thermal mass insulation
  • Hard ceramic hearth fitted as standard
  • Ventilated via top mounted ceramic chimney

Options (specify these at time of order)

  • Over-temperature protection (recommended to protect valuable contents & for unattended operation)

Laboratory Chamber Furnace - ELF

Laboratory Chamber Furnace - ELF



ELF 11/6ELF 11/14ELF 11/23
Max temp (°C)110011001100
Heat-up time (mins)284326
Max continuous operating temp (°C)100010001000
Dimensions:
Internal H x W x D (mm)
165 x 180 x 210210 x 220 x 310235 x 255 x 400
Dimensions:
External H x W x D (mm)
580 x 410 x 420630 x 450 x 520715 x 505 x 690
Temperature uniformity of ± 5°C within H x W x D (mm)125 x 140 x 140170 x 180 x 205195 x 215 x 305
Volume (litres)61423
Max power (W)200026005000
Holding power (W)90013001550
Thermocouple typeKKK
Weight (kg)243152

Please note
- Heat up rate is measured to 100°C below max, using an empty chamber
- Holding power is measured at continuous operating temperature
- External dimensions with door closed and including chimney
- The uniform volume is smaller than the total chamber volume

Content may be subject to modifications or corrections